lp-chrom@lipopharm.pl | 669 100 366









ECS05 Gradientowy System Analityczny

hplc
8 channel Scanning 200-800 nm 10 ml/min 40 MPa

High effective HPLC system containing UV-VIS PDA detector, column oven, analytical pump, gradient box and autosampler.

Detector with eight channels in range 200-800 nm and very lownoise level offers also high-speed scanning withspeed up to 100 Hz.

Communication of units is through RS232 or Ethernet(LAN). Additional PC A/D converter card is not required for this system. System contains autosampler and Clarity software. All units are supported in Clarity software from installation.

Main parts

ECDA2800 UV-VIS PDA DETECTOR
Analytical Diode Array Detector measures at eight wavelengths simultaneously or performs full scan with speed up to 100 Hz. 3D support in Clarity SW. Noise level ±5x10-6 AU. Four analog outputs. Two special mathematical channels are available. AD 05 analytical cell is included.
 
ECO2080 COLUMN OVEN WITH HEATING AND COOLING
This column oven allows precise temperature stabilization in the range 0-80°C. A big compartment room can accommodate up to 3 columns with length up to 25 cm. Peltier technology offers the possibility of heating and cooling independently from room temperature.
 
ECP2010 ANALYTICAL HPLC PUMP
This isocratic pump works with Gradient Box ECB2004B as a quaternary gradient pump. Flow rate range is 0.01-10 ml/min and pressure limit at 40 MPa. Unit software includes new learning algorithm for pulsation suppression and many testing and diagnostic functions.
 
ECB2004B GRADIENT BOX WITH DEGASSER
Box is suited for handling liquids in ECOM gradient systems. It accommodates a container for solvent bottles, built-in high-efficiency vacuum degasser and four-way gradient valve. Up to four solvent lines may be degassed simultaneously by one unit. Four-way gradient valve is controlled from the pump ECP2010.
 
AUTOSAMPLER L-3320
Automatic autosampler ensuring excellent accuracy and linearity of sample injection. Very low sample residue contamination, μl pick-up mode achieves zero sample loss. Automated selfchecking. Rheodyne valve ensures reliable operation.
 
CLARITY CHROMATOGRAPHY SOFTWARE
This system is fully controlled by Clarity Chromatography software, which is also a part of delivery. Clarity software complies CFR21GLP requirements.
 
 
SYSTEM CONTAINS 
 
Pcs Description P/N
1 ECDA2800 Analytical UV-VIS PDA Detector ANW00001
1 ECB2004B Gradient box with degasser ABA0000X
1 ECP2010 Analytical HPLC pump ACA0000X
1 ECO2080 Column Oven with heating and cooling AOA0000X
1 Autosampler L3320 for 2x48 vials ARA03320
1 Clarity – Station for 1 instrument AP000C50
1 Clarity module for HPLC control AP000A24
1 Clarity SW module for AS control AP000A26
1 Module for PDA data Clarity (SW) AP000A29
1 Accessories for analyt. system series 2000 ASA00180
Note: Column and computer are not included. PC system has to be equipped for handling 3 serial RS232 or 1 Ethernet port at least. Check also Clarity software requirements.
borosilicate glass

Kamush® Glass Blowing Laboratory

We offer custom desin of columns for FPLC and column chromatography. We also construct, modify and repair ather borosilicate glassware.


Details and realizations Columns for FPLC
chromatography software

LP-chrom® Chromatography Software

Modern software for older HPLC systems. Give a second life to your HPLC from Beckman, Knauer, Kontron, Gilson and others. Intuitive and easy to learn !

Get your copy See realizations